胸部撮影における重金属フィルタの物理特性と画像特性
坂本 弘巳東田 善治赤坂 勉豊福 不可依長 哲二鬼塚 昌彦加藤 豊幸丸石 博文朝原 正喜寺田 央早渕 尚文
著者情報
ジャーナル フリー

2001 年 21 巻 2 号 p. 95-105

詳細
抄録

We have investigated the performance characteristic of the heavy metal filters with higher atomic numbers by comparing their patient exposures, tube loadings, radiographic contrasts, and noise Wiener spectra with those of a combination of copper and aluminum filter which has been used widely in chest radiography.
Seven heavy metal filters were used for this study. As for a tungsten filter, two filters different in thickness were used. One is 0.05 mm thick, and the other 0.10 mm. The other metal filters were respectively combined with a tungsten filter with a thickness of 0.05 mm.
Among the all filters investigated, tube loading of tungsten filter with 0.05 mm thick is minimum. Tungsten with 0.1 mm thick and tungsten with 0.05 mm + barium show the larger advantages in patient exposure than those of the other filters.
It was found that the magnitude of patient exposure varied slightly wit h the difference of image receptor used. The use of heavy metal filters showed the small advantages in the patient exposure reduction (-20%) compared with that of conventional copper filter. The use of heavy metal filters showed increasing tube loading (-5times) compared with that of conventional copper filter. Noise Wiener spectra of heavy metal filters showed comparable with those of conventional copper filter.
In conclusion, the use of heavy metal filter combinations offer no significant advantages over optimal conventional filters.