Pulsatile Dialysis
M. MINESHIMAK. SAKAIY. NISHIMOTOK. KOBAYASHI
Author information
JOURNAL FREE ACCESS

1981 Volume 10 Issue 1 Pages 85-88

Details
Abstract

Pulse flow has been said to reduce overall transfer resistance. Therefore, pulse flow appliance to hemodialysis can improve total treatment efficiency.
In this report, pulsatile removal effects on K+, urea, and creatinine were investigated in vitro by using Asahi AM-06 dialyzer.
As a result, each mass transfer coefficient increased with 80-90 beats (min-1), higher amplitude, and lower Constant Flow Ratio, respectively.
In addition, same high efficiency will be expected for hemofiltration and hemoperfusion.