遠距離法と2つの飽和モデルによる小照射野のコリメータ散乱係数Scの推定
足立 宏美稲越 英機早川 岳英井上 富夫笠原 敏文五十嵐 智早川 英樹棚邊 哲史
著者情報
ジャーナル フリー

2008 年 64 巻 3 号 p. 306-315

詳細
抄録

To estimate the collimator scatter factor, Sc of small field sizes in which a mini-phantom cannot be fully included at the nominal treatment distance (NTD=100 cm), we measured the in-air output of 4 MV and 10 MV X-rays of a Varian’s Clinac 2100 C/D using a mini-phantom at NTD and at a long source-to-chamber distance (SCD=200 cm) with field-size defined at the isocenter down to 4.6×4.6cm2 and 2.3×2.3cm2, respectively. We then compared the fitted curve to the NTD dataset by a cumulative exponential distribution model with that by a cumulative Gaussian distribution (error function) model containing a zero-field extrapolated term derived from the long SCD dataset. The results showed that the zero-field extensions of two fitted curves coincided for a 4 MV X-ray, but a large discrepancy was seen between them for a 10 MV X-ray. Therefore, the Sc of small field sizes not measurable using a mini-phantom at the NTD can be well estimated by applying the cumulative exponential model to the NTD dataset in the case of a 4 MV X-ray beam filtrated with a cone-shaped flattener. However, to estimate the Sc of such small field sizes in the case of a 10 MV X-ray beam filtrated with a bell-shaped flattener, we consider it preferable to also measure in-air output at a long SCD and to apply the cumulative Gaussian model as described here.