日本放射線技術学会雑誌
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
ノート
Dual Source CTを用いた冠状動脈造影CTにおける使用基準の有用性 心電同期高速二重螺旋スキャンと螺旋スキャン,非螺旋スキャンの被ばく線量比較
庄司 友和樋口 壮典飯田 哲也稲垣 公俊加藤 洋
著者情報
ジャーナル フリー

2013 年 69 巻 3 号 p. 257-263

詳細
抄録

We believe that the patient exposure dose differs by heart rate for a coronary computed tomography angiography (CCTA), and we attempted to reduce patient exposure dose of the CCTA. Specifically, we made a clear difference of exposure dose between heart rate and optimal cardiac phase or some imaging methods. Next, we established criterion of the CCTA in our hospital, and usefulness was discussed. For examination methods, patients with a heart rate below 60 beats per minute (bpm) received a high-pitch spiral scan (Flash Spiral mode), those between 61 to 70 bpm received a step and shoot scan (SAS mode), and those of >70 bpm or an irregular heart rate received low-pitch spiral scan (Helical mode). The results of the clinical study showed that patient exposure dose reduced 87% in Flash Spiral mode (1.93±0.26 mSv) and 66% in SAS mode (4.88±1.24 mSv) compared with Helical mode (14.35±3.42 mSv). In our present study, we proved the usefulness of our criteria using CCTA. If the results of our present study become guidelines of CCTA users, we suggest that total patient exposure dose can be reduced.

著者関連情報
© 2013 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top