MTFフラクタル次元によるX線画像の画質評価
遠藤 駿李 鎔範蔡 篤儀原 武史
著者情報
ジャーナル フリー

2014 年 70 巻 3 号 p. 250-253

詳細
抄録

Objective evaluation, e.g., using the modulation transfer function (MTF) or the noise power spectrum (NPS) of an X-ray imaging system may not correctly correspond to the subject’s evaluation, e.g., when using the receiver operating characteristic (ROC). Moreover, it is difficult to measure MTF or NPS from clinical images. We therefore applied MTF fractal dimension to an X-ray imaging system. The MTF fractal dimension includes the frequency properties of the human eye in addition to quantitative complexity. In this study, we demonstrated a relationship between basic image quality and MTF fractal dimension using simulated and actual X-ray images.