画像再構成法の違いが低線量 CT perfusionの画質に及ぼす影響—模擬動態画像を用いた観察者実験—
大村 知己李 鎔範高橋 規之佐藤 祐一郎石田 嵩人豊嶋 英仁
著者情報
ジャーナル フリー

2015 年 71 巻 11 号 p. 1063-1069

詳細
抄録

CT perfusion (CTP) is obtained cerebrovascular circulation image for assessment of stroke patients; however, at the expense of increased radiation dose by dynamic scan. Iterative reconstruction (IR) method is possible to decrease image noise, it has the potential to reduce radiation dose. The purpose of this study is to assess the visual effect of IR method by using a digital perfusion phantom. The digital perfusion phantom was created by reconstructed filtered back projection (FBP) method and IR method CT images that had five exposure doses. Various exposure dose cerebral blood flow (CBF) images were derived from deconvolution algorithm. Contrast-to-noise ratio (CNR) and visual assessment were compared among the various exposure dose and each reconstructions. Result of low exposure dose with IR method showed, compared with FBP method, high CNR in severe ischemic area, and visual assessment was significantly improvement. IR method is useful for improving image quality of low-dose CTP.